Plunger Information B-4
v
V
Speed [Hz]
c
time in st
Calculation:
v = 50 Hz L = 14
V = 5800 Hz A = 3000 full steps
c = 500 Hz
Ramp Up Steps
A
Vv
L
1
22 2 2
2
5800 50
2 14 2500
481=
=
∗∗
= half steps
Ramp Down Steps
A
Vc
L
3
22 2 2
2
5800 500
2 14 2500
477=
=
∗∗
= half steps
If AA A
13
2+< (481 + 477 < 6000) then:
Ramp Up Time
t
V
v
L
1
5800 50
14 2500
16=
=
=. seconds
Ramp Down Time
t
V
c
L
3
5800 500
14 2500
15=
=
=. seconds
Constant Speed Steps
AAAA
213
2 2 3000 481 477 5042=−= = half steps