EasyManua.ls Logo

CAVRO XP 3000 - Page 38

CAVRO XP 3000
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Communication 3-2
NOTE When using the Pump:Link software to send commands to a device, use the
ASCII address values in Table 3-2
Table 3-2. Address Switch Settings in Hex (ASCII)
Switch
Setting
Single Device Dual Device Quad Device All Devices
Hex
Address
ASCII
Address
Hex
Address
ASCII
Address
Hex
Address
ASCII
Address
Address Value
to Send
0 31 1 41 A 51 Q 5F _
1322
2333 43C
3344
4355 45E55U
5366
6377 47G
7388
8399 49I 59Y
93A:
A3B; 4BK
B3C<
C3D= 4DM5D]
D3E>
E3F? 4FO
FSelf Test
The user can communicate with all pumps in the chain by using address “5Fh,” for
example to initialize all pumps at once. Then each pump can be controlled independently
by using addresses “31h” to “3Fh.”
NOTE Multiple address commands cannot be used to determine device status, nor
will they respond to Report commands. Each device must be queried
separately.

Table of Contents