Software Communication 3-11
6 Select the File menu, and choose Properties. The Properties dialog box appears.
7 Select the Settings tab, and enter or select these options:
Function, arrow, and Control keys act as:
– Select “Terminal keys”
Emulation:
– Select “Autodetect”
– Enter “500” in Backscroll buffer lines
Click the ASCII Setup button. The ASCII Setup dialog box appears.
8 Enter or select these options:
– Select “Send line ends with line feed”
– Select “Echo typed characters locally”
– Enter a Line delay of “0”
– Enter a Charater delay of “0”
– Select “Wrap lines that exceed terminal width”
9 Click OK to close the ASCII Setup dialog box, then click OK to close the Properties
dialog box.
10 Set the pump address to 0 or the appropriate address.
11 Set jumper JP1-2 to DT protocol (JP1-2 removed). Note that no jumper is needed for
microstep-enabled firmware. The communication protocol is detected automatically.
12 Power on the pump and initialize it by typing /1ZR and pressing Enter.
To run the pump, see the commands listed in “Using the XP 3000 Command Set” in
this chapter.