14. HOW TO USE THE WIRELESS PROFESSIONAL ON SPECIFIC
INDICATIONS

EN
WIRELESS PROFESSIONAL
14.10.1.4 Patient position
The patient is placed in the position most comfortable for him/her: prone position or seated facing a
medical table with a chest support.
14.10.1.5 Stimulation energy
The energy must be increased gradually until it causes clearly visible muscle twitches, which are required
to induce hyperaemia.
The mi-RANGE function makes it possible to work with certainty within a therapeutically eective range.
The stimulator prompts you to firstly increase the level of energy:
a beep sound accompanies the flashing “+” symbols.
When it detects that the muscles have started to pump, the “+” symbols will stop flashing. You are at the
minimum level of energy that provides therapeutic results.
If the stimulation is well tolerated by the patient, it is advised to increase the energy level slightly.
At the end of the treatment or during a break, a statistic showing the percentage of time spent in the
eective range will appear on the screen.