ECO 3 EXPERT Ultrasound Diagnostic System
Page 25 / 191
Peripherals
VGA output for external monitor
VIDEO output for B&W video printer
LAN port output
LAN for DICOM and image review station
USB 2.0 for flash drive
Foot switch
AC/DC adapter: MDS-060AAS19 B
Input: 100-250V~1.5-0.75A, 50-60Hz
Output: 19V 3.16A
DELTA ELECTRONICS, INC.
Battery Pack: BT-2500, 4400mAh, DONGGUAN POWER INC
3.5 Installation Procedures
NOTE: Please do not turn on the power switch until finishing all the installation and
necessary preparation.
3.5.1 Environment Condition
The system should be operated under the following environment.
3.5.1.1 Operation Environment Requirement
Ambient Temperature: 10 ℃~40
Relative Humidity: 30%75%RH
Atmospheric Pressure: 700hPa1060hPa
Strong radiation sources or powerful electromagnetic waves (e.g. electro-magnetic waves from
radio broadcasting) may result in image ghosting or noise. The system should be isolated from
such radiation sources or electromagnetic waves.
To prevent damage to the system, do not use in the following locations:
Exposed to direct sunlight
Subject to sudden changes in temperature
Dusty