ECO 3 EXPERT Ultrasound Diagnostic System
Page 78 / 191
6.11 Urology Measurement
Normally urology measurements are performed in B and 2B mode.
Freeze the required image under Urology examination, then press [Calc] to enter into Urology
measurement status.
Meas.
Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Kidney
Vol.(L/R)
ml
Refer to distance
Meas.
Formula: V
(π/6)×L×H×W
L: Kidney_L
H: Kidney_H
W: Kidney_w
Bladder
Vol.(L*W*H)
Vol.(Biplane)
ml
Refer to distance
Meas.
Formula: V
(π/6)×L×H×W
L: Bladder _L
H: Bladder _H
W: Bladder _w
Prostate
Volume
ml
Refer to distance
Meas.
Formula: V
(π/6)×L×H×W
L: Prostate _L
H: Prostate _H
W: Prostate _w
PPSA
ng/ml
Formula: PPSA
0.12×V
PSAD
ng/ml
Formula:
PSAD=SPSA/V
SPSA: input the SPSA
When create a new
Patient
Residual
Volume
ml
Refer to distance
Meas.
Formula: V
(π/6)×L×H×W
L: RVU _L
H: RVU _H
W: RVU _w
6.12 Cardiology Measurement
Freeze the required image under Cardiac examination, then press [Calc] to enter into Cardiac
status.
Meas.
Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Distance
cm
Refer to distance
Meas.

Table of Contents