ECO 3 EXPERT Ultrasound Diagnostic System
Page 69 / 191
Meas.
Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Area
area/circle
Area cm
2
Circle cm
Refer to Ellipse
and trace meas.
Ellipse and trace.
Volume
Volume
(1straight line)
ml
Refer to
distance Meas.
Formula: V =
(π/6)×D3
D means: Depth
Volume
(1ellipse)
ml
Refer to ellipse
meas.
Formula: V =
(π/6)×A×B2
A: Long Axis
B: Short Axis
Volume
(2 straight line)
ml
Refer to
distance Meas.
Formula: V =
(π/6)×D1×D22
D1: the longer distance D2:
the shorter distance
Volume
(3 straight line)
ml
Refer to
distance Meas.
Formula: V =
(π/6)×D1×D2×D
3
D1, D2, D3: Distance
Volume(1 straight
line 1 ellipse)
ml
Refer to
distance and
ellipse Meas.
Formula: V =
(π/6)×A×B×M
A: Long Axis
B: Short Axis
M: Distance
6.4 B General Measurement
Press [B], [2B] or [4B] to enter into B, 2B or 4B mode, and then press [Calc] to enter into
measurement status. Or press [Change] to choose the General measurement.
Meas.
Menu
Submenu
Unit
Meas. Method/
Meas. Formula
Comment
Distance
cm
Refer to
distance Meas.
Perimeter/Area<Ell
ipsr>
Area cm
2
Circle cm
Refer to Ellipse
and trace meas.
Press [UPDATE] to
change ellipse and trace.
Volume
(1Distance)
ml
Refer to
distance Meas.
Formula: V =
(π/6)×D3
D means: Depth