ECO 3 EXPERT Ultrasound Diagnostic System
Page 46 / 191
second line key of [Frequency] to decrease the frequency.
5.6.2 Dynamic
Dynamic range is used for adjusting the contrast resolution of B mode image and mode image,
compressing or enlarging the display range of gray scale.
At the real-time status, press the first line key of [Dynamic] to increase the Dynamic, and press the
second line key of [Dynamic] to decrease the Dynamic. The range is 30 to 90.
5.6.3 i-Image
In real status, press the corresponding key of [i-Image] to adjust, the range is 0~3.
5.6.4 Compound
In real status, press the corresponding key of [Compound] to adjust, and the range is 0~3.
The SRA can’t be edited after opening the compound.
5.6.5 PW
Display PW mode.
5.6.6 Gain
In real status, rotate [GAIN] knob to adjust the Gain, and the range is 0~255, the step is 5.
5.6.7 STC
STC curves can be used for adjusting gain compensation in different image depth.
Drag the slide of STC to adjust the value.
STC curve will disappear automatically 1 second later after stopping adjustment.
5.6.8 Depth
Press [DEPTH/FOCUS] selection knob until the indicator of [DEPTH] is lit, and then rotate the
knob to change the depth of image.
5.6.9 Focus Position
Press the [DEPTH/FOCUS] selection knob until the indicator of [FOCUS] is lit. Then turn the knob
to change Focus Position.