Setting Up the Generator
Setup, Tests, and Adjustment
Force FX Electrosurgical Generator C Service Manual 5-11
Connecting the CUSA Handpiece with CEM Nosecone
Connections for combined monopolar/ultrasonic surgery
Warning
Electric Shock Hazard
• Do not connect wet accessories to the generator.
• Ensure that all accessories and adapters are correctly connected and that no metal is exposed.
If you choose to use a monopolar
footswitch, you must use a only
Valleylab monopolar footswitch and
connect it to the Monopolar 1
Footswitch receptacle
Patient return electrode
CUSA
handpiece
with CEM
nosecone
CEM indicator
illuminates green
Connect to CUSA system