Performance Characteristics
3-2 Force FX Electrosurgical Generator C Service Manual
Performance Characteristics
General
Dimensions and Weight
Operating Parameters
Output configuration Isolated output
Cooling Natural convection; side and rear panel vents; fan
Display Eight digital seven-segment displays: 0.75" (1.9 cm)
each
Mounting Universal cart (UC8009), CUSA EXcel System, a
Valleylab Argon Gas Delivery Unit II, or any stable flat
surface
Width 14" (35.6 cm)
Depth 18" (45.7 cm)
Height 4 3/8" (11.1 cm)
Weight <18 lb (< 8.2 kg)
Ambient temperature
range
50° to 104° F (10° to 40° C)
Relative humidity 30% to 75%, noncondensing
Atmospheric pressure 700 to 1060 millibars
Warm-up time If transported or stored at temperatures outside the
operating temperature range, allow one hour for the
generator to reach room temperature before use.