Standards and IEC Classifications
Technical Specifications
Force FX Electrosurgical Generator C Service Manual 3-17
Electromagnetic Interference
When placed on or beneath an activated Valleylab electrosurgical generator, the Force FX
Electrosurgical Generator C operates without interference. The generator minimizes
electromagnetic interference to video equipment used in the operating room.
Electromagnetic Compatibility
The Force FX Electrosurgical Generator C complies with the appropriate IEC 60601-1-2
and IEC 60601-2-2 specifications regarding electromagnetic compatibility.
Notice
The Force FX Electrosurgical Generator C should not be used adjacent to or stacked with
equipment other than specified in the Force FX Electrosurgical Generator C
User Guide and
Service Manual. If adjacent or stacked use is necessary, the Force FX
Electrosurgical Generator C
should be observed to verify normal operation in the configuration in which it will be used.
The Force FX Electrosurgical Generator C intentionally applies RF energy for diagnosis or
treatment during activation. Observe other electronic medical equipment in the vicinity during the
Force FX Electrosurgical Generator C activation for any possible adverse electromagnetic effects.
Ensure adequate separation of electronic medical equipment based on observed reactions.