EasyManua.ls Logo

Covidien Force FX - Front Panel REM Module Replacement

Covidien Force FX
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Front Panel REM Module Replacement
Replacement Procedures
Force FX Electrosurgical Generator C Service Manual 7-11
Front Panel REM Module Replacement
1. Remove the front panel assembly (observe all warnings and cautions). Refer to Front
Panel Replacement on page 7-9.
2. Remove the two screws that secure the REM module to the front panel and remove
the bracket that covers the REM module. Set the bracket and screws aside for
reinstallation.
3. Replace the existing module with a new REM module.
4. Position the bracket over the new REM module and install the two screws that secure
the module to the front panel.
5. Reinstall the front panel assembly. Refer to Front Panel Replacement on page 7-9.

Table of Contents

Related product manuals