Footswitch Board Replacement
7-8 Force FX Electrosurgical Generator C Service Manual
Footswitch Board Replacement
1. Remove the Power Supply/RF board and attached heat sinks (observe all warnings
and cautions). Refer to Power Supply/RF Board Replacement on page 7-26.
2. Remove the Footswitch board assembly.
a. On the rear panel, remove the six screws that secure the footswitch receptacles to
the rear panel.
b. Inside the rear panel, remove the three screws that secure the Footswitch board
assembly to the rear panel standoffs.
3. Install the new Footswitch board assembly.
a. Position the assembly inside the rear panel and install the three screws that secure
the assembly to the rear panel.
b. On the rear panel, install the six screws that secure the footswitch receptacles to
the rear panel.
4. Reinstall the Power Supply/RF board and attached heat sinks. Refer to Power Supply/
RF Board Replacement on page 7-26.