Standards and IEC Classifications
3-18 Force FX Electrosurgical Generator C Service Manual
Cables Used for the Force FX EMC Test
Force FX EMC Test—Cables Used
Model Function Length Location
207002060 AC cable 15 ft. Rear panel
Multicontact Part Number:
15.2013-46020
Equipotential cable 15 ft. Rear panel
E2450HDBNSB Two-button pencil 15 ft. Front panel mono 1
E2450HDBNSB Two-button pencil 15 ft. Front panel mono 2
E7507DB Adult REM pad 15 ft. Front panel REM
E0020V / E4073CT Foot switching bipolar
cord / bayonet forceps
15 ft. Front panel bipolar
E6019 Dome bipolar footswitch 15 ft. Rear panel bipolar
E6008B Monopolar footswitch 15 ft. Rear panel mono 1
E6008B Monopolar footswitch 15 ft. Rear panel mono 2
VLCM Bipolar current monitor 24 ft. Rear panel
EA3730 EKG blanking cable 5 ft. Rear panel
Warning: The use of accessories, other than specified in the Force FX-C User’s Guide and Service
Manual, may result in increased emissions or decreased immunity of the Force FX-C generator.
Only the use of Covidien-branded monopolar cables and accessories, bipolar cables and
accessories, and footswitches are approved for use with the Force FX generator. Use of non-
Covidien-branded cables and accessories have not been assessed and therefore do not comply
with the safety and EMC requirements of IEC 60601-1, IEC 60601-1-2, and/ or IEC 60601-2-2.