EasyManua.ls Logo

Covidien Force FX - Page 149

Covidien Force FX
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating the Generator
Setup, Tests, and Adjustment
Force FX Electrosurgical Generator C Service Manual 5-41
c. Stop activation. Verify that the Coag display changed to 120.
d. Press the coag pedal again and, while activating the generator, press the Coag up
() or down () arrow buttons until the voltmeter reading is equivalent to
489 ± 6 mA
rms
. Ignore changes to the Coag display.
e. Stop activation. Verify that the Coag display changed to 10.
6. To save the ECON calibration values, press the Bipolar up arrow () button.
To leave Calibration mode, refer to Exiting Calibration Mode on page 5-28.

Table of Contents

Related product manuals