Correcting T_ON ASIC Malfunctions
Troubleshooting
Force FX Electrosurgical Generator C Service Manual 6-23
d. Connect the Display board ribbon cable to the Control board and lock the
connector.
5. Calibrate the generator. Refer to Calibrating the Generator on page 5-26 for
instructions.
If the alarm number reappears, record the number and call the Covidien Service
Center.
6. To install the cover, position the cover above the chassis and slide it down. Install the
five screws that secure the cover to the chassis.