FIELD REPAIR/SERVICE MICRO-TEMP
®
LT, Model 749
OPERATION AND TECHNICAL MANUAL
Page 60 of 75
H. Audible indicator
sounds.
Temperature readout
indicates 111˚F / 44˚C or
higher.
Unit has reached high limit.
Turn unit off and allow unit
to cool. Service required.
Restore power to the unit.
Place the unit on a level
surface.
Temp >1˚C over set
point.
Check pad connections.
Wait a few minutes for
water temp to decrease.
Water level sensor fault.
Turn OFF unit and make
repairs based on the
display message. Refer to
Section (3-5-B).
Water Temperature
Sensor defect.
I. Red “Power
Failure/Caution” LED
on
Check pad connections.
Wait a few minutes for
water temperature to
decrease. If the display
alternates between flashing
“HL” and the actual water
temperature, turn unit off,
allow unit to cool and then
service required.