MICRO-TEMP LT
Operation and Technical Manual
Model 749 Localized Therapy Units
Cincinnati Sub-Zero Products, LLC. 12011 Mosteller Road Cincinnati, Ohio 45241, U.S.A.
www.cszmedical.com

Table of Contents

Questions and Answers

  • R
    Randy RiddleSep 10, 2025
    What to do if the CSZ Medical Equipment audible alarm sounds?
    • M
      Melissa ReidSep 12, 2025
      If the audible alarm sounds on your CSZ Medical Equipment, it could be due to several reasons. The temperature readout might indicate 111°F / 44°C or higher, meaning the unit has reached its high limit; turn it off and let it cool. The unit might be tilted, so ensure it's placed on a level surface. Also, check pad connections if the temperature is more than 1°C over the set point.
  • Y
    youngcarlaSep 9, 2025
    Why is the CSZ MICRO-TEMP LT 749 unit not turning on?
    • V
      Victoria CollinsSep 9, 2025
      The CSZ Medical Equipment unit might not turn on due to several reasons. It could be that the unit is not plugged into the proper voltage supply. Ensure it is connected to the correct voltage. Another cause might be a blown fuse, in which case you should replace the fuse. Also, check for improper connections to the control board and correct them.
  • M
    Marc WrightSep 8, 2025
    What should I do if the control board fails on my CSZ Medical Equipment?
    • B
      Benjamin HaynesSep 8, 2025
      To address a control board failure in your CSZ Medical Equipment, turn off the unit and replace the control board.
  • A
    amanda70Aug 22, 2025
    How to fix a CSZ MICRO-TEMP LT 749 Medical Equipment unit that will not turn on?
    • R
      Ricky WarrenAug 23, 2025
      If your CSZ Medical Equipment unit isn't turning on, there are a few potential causes. First, ensure the unit is plugged into the correct voltage supply, referring to Section (7) for electrical specifications. If it is, check the fuse; a blown fuse will prevent the unit from powering on, so replace it according to Section (5-9). Lastly, verify that the connections to the control board are properly configured, consulting Section (5-2.3) for guidance.
  • C
    Christopher SharpSep 5, 2025
    Why does my CSZ MICRO-TEMP LT 749 Medical Equipment heat, but the pump will not run?
    • S
      sheliadavisSep 5, 2025
      If your CSZ Medical Equipment heats but the pump doesn't run, it could be due to a defective pump, which would need replacement. Another possibility is improper connections to the control board; ensure these connections are correctly configured.
  • M
    Mario KennedySep 4, 2025
    How to resolve a water level sensor fault in CSZ Medical Equipment?
    • R
      Ryan WellsSep 4, 2025
      To address a water level sensor fault in your CSZ Medical Equipment, turn off the unit. The water level sensor may need replacement or repair.
  • C
    Chelsea ThomasSep 3, 2025
    What causes ground continuity to exceed 0.2 ohm in CSZ Medical Equipment?
    • A
      allisonharrisonSep 3, 2025
      If the ground continuity exceeds 0.2 ohm in your CSZ Medical Equipment, check for loose or disconnected ground (green) wires and tighten or reconnect them. The connector on the ground wire may be defective and need replacement. Also, the power cord itself could be defective and require replacement.
  • J
    jacobcrawfordSep 2, 2025
    What to do if my CSZ MICRO-TEMP LT 749 shows a high temperature?
    • C
      Chelsea WilliamsSep 2, 2025
      If your CSZ Medical Equipment is showing a high temperature, check the pad connections. You can also wait a few minutes for the water temperature to decrease, or turn off the unit and allow the water to cool down.
  • P
    Phillip CampbellSep 1, 2025
    Why will the display not illuminate on my CSZ MICRO-TEMP LT 749?
    • J
      Joshua SmallSep 1, 2025
      If the display on your CSZ Medical Equipment isn't illuminating, the white ribbon cable connector between the display board and control board might be disconnected or improperly connected. Ensure the connections are correct. Alternatively, the control panel and/or display board could be defective, requiring replacement.
  • N
    nancy38Aug 31, 2025
    What causes leakage current to exceed limits in CSZ MICRO-TEMP LT 749?
    • G
      George WilliamsAug 31, 2025
      If the leakage current exceeds the specified limits on your CSZ Medical Equipment, it could be due to a defective power supply, which would need replacement. It may also be caused by a defective heater or pump, each requiring replacement.

Summary

Safety Warnings

General Warnings

Lists critical warnings regarding patient safety, equipment misuse, and potential hazards.

Operational Cautions

Cautions for Safe Operation

Provides cautions for proper unit setup, usage, and maintenance to prevent damage.

System Introduction

General Safety Precautions

General safety precautions for using the localized heat therapy system.

Temperature Safety Controls

Temperature Safety Control System

Explains the safety features designed to monitor and control water temperature.

System Preparation

First Time Set-Up;System Test Routine

Outlines steps for initial setup and performing a system test.

Setup Details

Completing System Test Routine

Describes the steps to complete a system test routine for the unit.

Unit and Patient Precautions

Unit and Patient Related Precautions

Lists precautions related to the unit's operation and patient care.

Operating the System

System Operation

Operating the System

Provides step-by-step instructions for operating the system.

General Maintenance

Maintenance Requirements

Outlines general maintenance requirements for regular system operation.

High Limit and Low Water Checks

High Limit Safety Check

Explains how to perform a check on the high limit safety feature.

Field Repair and Service Introduction

Accessing Internal Components

Access to the Interior

Explains how to safely access the internal parts of the unit.

Enclosure Separation and Reconnection

Separating Front and Rear Enclosures

Provides steps for separating the front and rear enclosures of the unit.

Reconnecting Front and Rear Enclosures

Details the procedure for reconnecting the unit's front and rear enclosures.

Cable Disconnection Procedures

Disconnecting Cables from Control Board

Explains how to disconnect cables from the control board.

Disconnecting Cables from Display Board

Details how to disconnect cables from the display board.

Heater Replacement

Replacement of the Heater

Provides instructions for replacing the unit's heater.

Pump Replacement

Reservoir Replacement

Tubing and Sensor Replacement

Power Switch and Fuse Replacement

Board and Panel Replacement

Replacement of Control;Display Board

Instructions for replacing the control board and display board.

Replacement of Membrane Panel

Instructions for replacing the membrane control panel.

Troubleshooting Guide

Troubleshooting Guide

Provides a guide to diagnose and resolve common unit issues.

System Specifications

CSZ MICRO-TEMP LT 749 Specifications

General IconGeneral
Temperature Stability±0.1°C
Heating Capacity1000 Watts
Communication InterfaceRS-232
Accuracy±0.2°C
Resolution0.1°C
Pump Flow Rate10 liters per minute
Controller TypeMicroprocessor-based
DisplayLCD
Power Supply120V/60Hz or 230V/50Hz
Probe TypeRTD