EXCEL V OPERATOR MANUAL
D1025, REV H, 10/17
System does not turn on. The touchscreen does not illuminate.
Treatment beam is not present, although aiming beam operates properly.
Treatment and aiming beams are not present. Touchscreen displays and indi-
cators are normal.
Probable Cause The system is not plugged in.
Suggestion Place the system main power circuit breaker in the OFF
(down) position, insert the system electrical power cord plug
into the appropriate outlet, and place the system main power
circuit breaker in the ON (up) position. Turn the keyswitch to
the START position.
Probable Cause The building power (main electrical service) is turned off.
Suggestion Turn on the building power or reset the room circuit breaker.
Probable Cause The system main power circuit breaker is in the OFF (down)
position.
Suggestion Place the system main power circuit breaker in the ON (up)
position.
Probable Cause The electrical outlet is defective.
Suggestion Use another outlet or have the outlet professionally tested
and, if appropriate, repaired
Probable Cause The remote interlock plug is not connected to the system
console.
Suggestion Insert the remote interlock plug into the interlock receptacle,
then turn the keyswitch to the START position.
Probable Cause There is an internal laser system failure.
Suggestion Contact Cutera Service (refer to Appendix D).
Probable Cause The system is in STANDBY mode.
Suggestion Place the system in READY mode.
Probable Cause There is an internal laser system failure.
Suggestion Contact Cutera Service (refer to Appendix D).