EasyManua.ls Logo

Cutera ExcelV - Physical Parameters; Electrical Requirements; Environmental Requirements; Maximum Allowable Temperature of Accessible Parts

Cutera ExcelV
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
63
EXCEL V OPERATOR MANUAL
D1025, REV H, 10/17
Physical Parameters
Size (W X D X H): 16.1 in. X 24 in. X 34.8 in. (.41 m x .61 m x .88 m)
Weight: 185 lbs. (84.1 kg.)
Operating length of umbilical: 9 ft. (2.7 m)
Minimum bend radius of umbilical: Storage 5 in. (126 mm)
Momentary 2 in (50 mm)
Electrical Requirements
Voltage, Current 100-120 VAC, 20 A or 200-240 VAC, 16 A
Frequency 50/60 Hz
Environmental Requirements
Temperature Operating C to 30º C
Non-operating -10º C to 50º C, coolant drained
Humidity 0 to 90%, non-condensing
Max. Heat Dissipation 4,000 Watts
Max. Altitude Operating 15,000 ft.
Non-operating 50,000 ft.
Atmospheric Pressure Operating: 60 to 110 kPa
Maximum Allowable Temperature of Accessible Parts
Parts User/Patient Can Touch Material Duration Limits
Enclosure Plastic > 1 min 48º C
Touchscreen Plastic > 1 min 48º C
Coolview Handpiece Window Sapphire > 1 min 20º C
CoolView Handpiece Treatment Guide Metal > 1 min 20º C

Table of Contents

Related product manuals