Electromagnetic Compatibility
The Excel V design complies with IEC 60601-1-2 (3rd edition) requirements for electromagnetic
compatibility (EMC) with other devices. Like other electrical medical equipment, the Excel V requires
special precautions to ensure EMC with other electrical medical devices and must be installed and
operated according to the EMC information provided in this manual.
CAUTION
Portable and mobile RF communications equipment may affect the
normal function of the Excel V.
WARNING
Do not use cables or accessories other than those provided with the
Excel V, as this may result in increased electromagnetic emissions
or decreased immunity to such emissions.
WARNING
If the Excel V is used adjacent to or stacked with other equipment,
observe and verify normal operation of the Excel V in the
configuration in which it will be used prior to using it in a surgical
procedure. Consult the tables below for guidance in placing the
Excel V.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The Excel V is intended for use in the electromagnetic environment specified below. The customer or the user
of the Excel V should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment: Guidance
RF emissions CISPR 11 Group 1 The Excel V uses RF energy only for its internal function;
therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B The Excel V is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply net-
work that supplies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations/flicker
emissions IEC61000-3-3
Complies