62
EXCEL V OPERATOR MANUAL
D1025, REV H, 10/17
System Specifications
Specifications subject to change without notice.
Treatment Beam
Type High Power, Frequency-doubled KTP High Power, Long Pulse Nd:YAG
Wavelength 532 nm 1064 nm
Output 12 J Max. 100 J Max.
Fluence
1.8 to 42 J/cm
2
CoolView: 2 to 300 J/cm
2
Genesis V: 4 to 7 J/cm
2
Pulse Duration 1.5 to 40 ms CoolView: 5 to 60 ms
Genesis V: 0.3 ms
Repetition Rate 0.0 Hz (single shot) to 4.0 Hz CoolView: 0 Hz (single shot) to 2.0 Hz
Genesis V: 0 Hz (single shot) to 10.0 Hz
Treatment Spot 2 to 12 mm circular spot CoolView: 2 to 12 mm circular spot
Genesis V: 8 mm circular spot
Delivery System CoolView handpiece with
user-detachable umbilical
CoolView and Genesis V handpieces with
user-detachable umbilical
Working Distance n/a CoolView: n/a
Genesis V: 3 to 4 cm
Epidermal Cooling User-adjustable contact sapphire cooling CoolView: User-adjustable contact
sapphire cooling
Genesis V: n/a
System Cooling Self-contained water to air heat exchanger
Aiming Beam 650 nm