38
Dakewe (Shenzhen) Medical Equipment Co., Ltd
timing. In order to prevent forgetting to clean, there will be a cleaning machine
reminder. The default is from the start of the timing to 30 minutes when there is no
cleaning, there will be a sound prompt, and the main interface will prompt Need to run a
cleaning program.
2. Boot preferences 2
After finishing the preference setting page 1, click "Next" to enter the "Preferences 2"
interfaceto set. To go back to the previous setting page, click "Previous".
Preferences 2
1Choose whether to turn it on according to the actual use. The washing and
drying function is turned on by default. Turning on this function will keep the tissue
cylinder dry and odorless. Turning off this function will reduce the cleaning procedure
time by 15 minutes;
(2) Function of melting residual wax in tissue basket Before running the first
step of the standard cleaning procedure, pump paraffin from W1 to soak in the tissue
processing tank to melt the residual wax attached to the tissue basket and return it to W1,
which reduces the consumption of paraffin and the loss of cleaning reagents. The
melting is turned off by default. Tissue basket residual wax function;
(3) Preset delay end time When you select a delay in the running program, the
end time will be set to this preset time (the default is 8 am), and you can modify this
time according to your habits.