6.2.4 Paraffin purification
Click "Paraffin Purification Setting" to enter the paraffin purification setting interface. After the
paraffin purification function is enabled, the system will enter the intelligent mode by default. In
smart mode, you can set the wind speed and duration. Every day after the cleaning process is
finished, the paraffin purification process will be automatically run according to the set value
(only once). You can also set the mode to manual. In manual mode, you can set the wind speed
(low, middle high three levels), purification duration (15min, 30min, 45min, 60min), purification
start time, whether to repeat every day (purification once every day at the start time), start mode
(immediate start or delay start); Click Start to start the program.