76   A-Class 
 
Step
Action  Displayed Message  Comment 
G 
Observe the calibration 
result. 
CALIBRATION COMPLETE 
Calibration was 
successful. 
 
;
 If ‘Warning Low 
Backing’ is displayed, 
calibration was 
successful (for 
possible messages 
see Section 5.1). 
 
H 
Press the ESC Key 
repeatedly to exit menu 
mode. 
CALIBRATION COMPLETE 
 
Followed by... 
 
READY 
When calibrating gap or 
reflective media, press 
and hold the FEED Key 
until at least one label is 
output. 
 
The printer is now ready 
for use.
  
 
 
 
 
 
 
5.2.2  Advanced Entry  
 
;
 
Advanced Entry will override all previous calibration settings and should only be 
used when Standard Calibration attempts fail. Ensure that the correct SENSOR 
TYPE is selected (see Section 4.2.1) before staring this procedure. 
 
Advanced Entry is an alternate calibration method for special-case media types, where 
sensor readings are taken using different sampling algorithms and from a list of these 
readings the best algorithm is selected for manual entry into the database. Follow the steps 
below to perform an Advanced Entry Calibration:   
 
Step
Action  Displayed Message  Comment 
A  Turn ON the printer.  
CANNOT CALIBRATE 
 
-or- 
 
UNCALIBRATED  
Wait briefly for the printer 
to initialize.  
 
(continued)