Maintenance and Adjustments 
2-11
2.7.2  Manual Calibration  
 
If media sensing problems continue after Auto Calibration, use the procedure below.   
 
1. Hold the 
 button while powering ON the printer and until the PAUSED Light turns OFF, then 
release the button.   
 
2.  Press and hold the 
 button until the PAUSED Light turns ON then release the button. 
 
3.  Place media (with backing material attached, if any) over the Media Sensor then close the 
Printhead Assembly and press the 
 button. The FAULT Light will flash as the “paper” is analyzed. 
 
4.  Position only the backing material (or the black mark if using reflective media) over the Media 
Sensor then close the Printhead Assembly and press the 
 button. The FAULT Light will flash as 
the “TOF” is analyzed. 
 
5.  Remove the media then close the Printhead Assembly and press the 
 button. The FAULT Light 
will flash as the “empty” is analyzed. 
 
6.  Simultaneously and briefly press the 
 + 
 buttons to temporarily accept the settings and exit 
calibration. One of the following indicators will flash five times to denote the calibration result:  
 
PAUSED Light = Successful (proceed to Step 6); or,   
 
FAULT Light = Unsuccessful (return to Step 1 to retry the procedure).  
 
  (Test the calibration: Use the 
 button to feed labels and use the   button to print test labels.) 
 
;
 Note:  To discard the changes and revert back to the previous calibration settings, turn OFF 
the printer before executing Step 7. 
 
7. Press the 
+
 buttons simultaneously and briefly. Wait until the PAUSED Light goes OFF.