Passport 2®/Passport 2 LT™ Service Manual 0070-10-0441 6 - 11
Calibration Procedure Diagnostics
6. Adjust R81 potentiometer and verify the linearity accordingly. See figure below.
FIGURE 6-13
Specification: = 150mmHg +/- 0.1%
7. Reassemble unit and verify the following static pressure points to ensure proper
calibration.
Specification: 30 mmHg, 100 mmHg, 150 mmHg, 190 mmHg and 250
mmHg. +/-3 mmHg.
150 mmHg
Digital Mercury Manometer
Bulb
693cc
Test Chamber
P/N 0138-00-0001-01
T - Fitting
Overpressure Voltage Setpoint
Static Pressure Calibration
Motor Pump Test
Leak Test
Dump Test
Bleed Rate Test
Overpressure Test
Abort Test
Previous Menu
Select to return to previous menu.
Transducer: 150.0 mmHg
Passport
2
Static Calibration Setup
Bulb
Valve
VR 81
NIBP Module
Manually
Squeeze
Bulb
Static Pressure Calibration