Passport 2®/Passport 2 LT™ Service Manual 0070-10-0441 6 - 17
Calibration Procedure Verification
6.5 Verification
6.5.1 Initial Set-up
1. Using a patient simulator, connect the ECG, IBP1, IBP2 and temperature cables to the
left side connector panel. Set the ECG simulator for 60 bpms, 1mv QRS signal.
2. Set up the Passport 2 as follows:
Patient Menu - Adult mode (patient size)
Monitor Setup -
a. Display Setup - 3 Waveforms / 6 Waveforms (12Leads)
b. ECG Speed - 25 mm/sec
c. IBP Speed - 25mm/sec (optional)
d. Respiration / Gas Speed - 12.5 mm/sec
Print Setup -
a. Waveform 1 - ECG 1
b. Waveform 2 - ECG 2
c. Select Printer - Local
Parameters -
a. ECG
ECG1 - II
ECG 2 - I
ECG 3 - III
ECG 1 through 6 Size - 1cm/mV
(12 Lead) Page 1
ECG 1 - I
ECG 2 - II
ECG 3 - III
ECG 4 - AVR
ECG 5 - AVL
ECG 6 - AVF
(12 Lead) Page 2
ECG 1 - V1
ECG 2 - V2
ECG 3 - V3
ECG 4 - V4
ECG 5 - V5
ECG 6 - V6