Patient Connector Board H.P. Theory of Operation
2 - 32 0070-10-0441 Passport 2®/Passport 2 LT™ Service Manual
2.6.7 0670-00-0680-01
This module is responsible for acquisition of most of the patient-safety-isolated signals,
namely the ECG, Temperature and two channels of Invasive Blood Pressure connectors IBP1
and IBP2. The Connector board collects all of the individual patient cables into a single
ribbon cable, then passes the signals to the main board.
2.6.8 0670-00-0680-02
This module is responsible for acquisition of the patient-safety signals, namely the ECG and
the Temperature. This module is without the dual 12- pin connectors for the IBP. The
Connector board collects all of the individual patient cables into a single ribbon cable, then
passes the signals to the main board.