EasyManua.ls Logo

Delta Elektronika SM15K User Manual

Delta Elektronika SM15K
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
Ethernet & Sequencer Programming SM15K
12 / 44 DELTA ELEKTRONIKA B.V. rev. Aug. 2019
The result will be in Watt Hour (Wh), scientific notation.
If the instrument is not enabled, the result will be zero.
To read the total Watt Hour for the negative power:
Syntax: MEASure:INStrument<sp>WH,NEG,TOTAL?<term>
The result will be in Watt Hour (Ah), scientific notation.
If the instrument is not enabled, the result will be zero.
To read the minimum positive power during the time the instrument is enabled:
Syntax: MEASure:INStrument<sp>WH,POS,PMIN?<term>
The result will be in Watts. If the instrument is not enabled, the result will be zero.
To read the maximum positive power during the time the instrument is enabled:
Syntax: MEASure:INStrument<sp>WH,POS,PMAX?<term>
The result will be in Watts. If the instrument is not enabled, the result will be zero.
To read the minimum negative power during the time the instrument is enabled:
Syntax: MEASure:INStrument<sp>WH,NEG,PMIN?<term>
The result will be in Watts. If the instrument is not enabled, the result will be zero.
To read the maximum negative power during the time the instrument is enabled:
Syntax: MEASure:INStrument<sp>WH:NEG,PMAX?<term>
The result will be in Watts. If the instrument is not enabled, the result will be zero.
Measure temperature
To read the highest internal temperature of the power supply:
Syntax: MEASure:TEMperature?<term>
The answer is in degrees Celsius..
5.4 Calibrate Subsystem
5.4.1 Introduction
The calibration of the power supply is done during production. However, periodical check and calibration is
recommended. At power-on, the calibration settings are restored.
There are four parameters to calibrate. Two related to the voltage measurement and two related to the
current measurement.
For proper calibration it is recommended to use the following order (an SM500-CP-90 is used as example) :
- Set output voltage of the power supply to e.g. 1% of the maximum output voltage.
SOURce:VOLtage<sp>5.00<term>
- Calibrate the measure voltage offset, so the result of the command MEASure:VOLtage? is as
close as possible to 5.00 V.
- Set the output voltage to maximum
SOURce:VOLtage<sp>500<term>
- Calibrate the measure voltage gain, so the result of the command MEASure:VOLtage? is as
close as possible to the actual output voltage, 500 V.
Same principle is used for the current calibration.
Default settings for the offsets are 0, and default settings for the gains are 1. The resolution of both settings
and readings are 6 digits.
Notice: Both gain and offset changes influence the actual output parameter. After changing offset, check if
the gain has to change. And visa versa. To get a very accurate result, redo the calibration a few times. To
save the calibration settings to the non-volatile memory, refer to section 5.1
For MEASURE offset calibration, use the equation:
For MEASURE gain calibration, use the equation:

Table of Contents

Other manuals for Delta Elektronika SM15K

Question and Answer IconNeed help?

Do you have a question about the Delta Elektronika SM15K and is the answer not in the manual?

Delta Elektronika SM15K Specifications

General IconGeneral
Output Power15 kW
ProtectionsOvervoltage, Overcurrent, Overtemperature, Short circuit
CoolingForced air
Input Voltage3x400 Vac
Input Frequency45-65 Hz
Operating Temperature0 to +50 °C
Storage Temperature-40 to +70 °C

Summary

In General

Features

Describes the remote Ethernet programming function for automated systems, parameter control, and data acquisition.

Installation

Infrastructure

Covers connecting the Ethernet interface via RJ45 to a network or PC, mentioning Auto-MDIX.

Communication

Settings

Details the two key settings for communication: IP address (default DHCP) and fixed port number 8462.

Command Description

General Instructions

Covers general commands like *IDN?, *PUD, *SAV, *CLS, and *RST for identification, data storage, and reset.

Source Subsystem

Details commands for setting and reading output voltage, current, and power, including maximum values and step sizes.

Measure Subsystem

Explains commands to measure actual output voltage, current, and power, including internal instruments for Ah/Wh.

Sequencer

Introduction

Introduces the SEQUENCER subsystem, capable of 25 free programmable sequences with 2000 steps each.

Commands

Details sequencer commands categorized into Settings, Jumps, Arithmetic, and Miscellaneous operations.

Sequence control by commands

Explains how to control sequences via commands, including catalog, creation, upload, delete, start, pause, and stop.

Related product manuals