Basic Weighing Function
Preparation
§ Switch on the balance: Press
§ Tare the balance, if necessary:
Press t
$ If necessary, change the configura-
tion settings:
See the chapter entitled
“Configuration”
$ If desired, load the factory settings:
See the chapter entitled
“Configuration”
Additional Functions
$ Switching off the balance:
Press
$ Balance in standby mode:
the current time is displayed
Features
– Taring the balance
– Printing weights
Using Verified Balances as Legal
Measuring Instruments in the EU*:
The type-approval certificate for
verification applies only to non-auto-
matic weighing instruments.
For automatic operation with or with-
out auxiliary measuring devices, you
must comply with the regulations
applicable to the place of installation.
§ Before using the balance as a legal
measuring instrument, calibrate
and adjust it at the place of use
using the built-in motorized cali-
bration weight; for details, see
“Calibration/Adjustment” in this
chapter.
$ The temperature range (°C)
indicated on the verification label
must not be exceeded during
operation.
Example:
BD ED 200
K +10°C to +30°C
* Including the Signatories of
the Agreement on the Eruopean
Economic Area