34
Animal Weighing/Averaging
Display symbol: V
Purpose
Use this program to determine the
weights of unstable samples (e.g., live
animals) or to determine weights
under unstable ambient conditions.
With this program, the balance calcu-
lates the weight as the average of a
defined number of individual weighing
operations (also referred to as “sub-
weighing operations”).
Preparation
§ Select the Animal weighing
application in the menu:
see “Configuration.”
§ Set the following parameters:
Applic. Application program
AnimalW. Animal weighing
activty. Animal activity
Calm Stable readout
o Active Unstable
V.Active Very unstable readout
Start
manual
o Autom. Automatic
o = Factory setting
Changing the Number of Subweigh-
ing Operations
Activate function:
Press the key
Select the desired number of
measurement (1 to 100):
In increments of 1:
Press the key briefly
In increments of 10:
Press and hold the key.
The selected number of measure-
ments is stored in battery-backed
memory.
Printout: Animal weighing
mDef 20 : Number of sub-
weighing operations
x-Net + 410.1 g : Calculated average