Counting
Display symbol: Z
Purpose
With the Counting program you can
determine the number of parts that
each have approximately equal weight.
To do this, a known number of parts
(the reference sample quantity) is
weighed first, and the individual
piece weight (reference weight) is
calculated from this result.
Thus the number of parts subse-
quently placed on the balance can be
determined from their weight.
Preparation
§ Select the Counting application in
the menu:
see Configuration.
§ Set the following parameters:
Applic. Application program
Count.
Resolut. Resolution
o Disp.Acc. Display accuracy
10-fold 10-fold higher
Ref.Updt. Autom. ref. sample
updating
o Off Display accuracy
Autom. Automatic
o = Factory setting
Changing the Reference Sample
Quantity
Activate function:
Press the key
Select the desired reference sample
quantity (1 to 100):
In increments of 1: Press the key
briefly
In increments of 10:
Press and hold the key.
The quantity is stored in battery-
backed memory.
Reference Sample Updating
Automatic reference sample updating
optimizes the counting accuracy.
You can activate or deactivate this
function in the menu.
Automatic reference sample updating
is performed when the requirements,
including the specified stability crite-
rion, have been met.
The abbreviation opt, for optimiz-
ing, is displayed briefly with the
new reference sample quantity.
Printout: Counting
nRef + 10 : Reference sample
quantity
wRef + 21.14 g : Reference weight
Qnt + 500 pcs : Calculated quantity