Example : Counting parts of equal weight
Parameter: Applic. - Count. (menu code 2. 3.)
Step Key Display/Data output
(or instruction)
1. Place empty container +
22.6 g
on the balance
2. Tare the balance t
0.0 g
3. Add reference sample quantity
to container (in this example:
20 pcs)
4. Changing the reference
ref 10
pcs
sample quantity:
5. Select reference sample quantity: Repeatedly:
In increments of 1 (1, 2, 3, etc. to 100) Press briefly
In increments of 10 (10, 20, etc. to 100) press
and hold
ref 20
pcs
6. Confirm selected reference + 20
pcs
*
sample quantity and start nRef 20 pcs
application. The current reference wRef 1.07 g
weight remains stored until a new
reference weight remains stored
until a new reference is set or
the power supply is interrupted
7. Add desired number of pieces. +
500
pcs
8. If desired, print quantity Qnt + 500 pcs
9. Toggle display between 1.07 g
G
*
mean piece weight, weight, Repeatedly: + 535.0 g
*
quantity + 500
pcs
*
10. Unload the balance 21
pcs
*
11. Repeat as needed, starting from
Step 7
12. Delete reference value
0.0 g
CLEAR
SELECT
PRINT
ENTER
SELECT
SELECT
SELECT