45
Line format for printout: include
data ID codes (22 characters;
factory setting):
Setup: Prnt.Out: Format: 22 Char.
(menu code 1. 6. 6. 2)
Formats for time:
setup: prnt.0ut: time: 24 h
(menu code 1. 6. 8. 1) or
setup: prnt.oUt: time: 12 h
(menu code 1. 6. 8. 2)
Formats for date:
setup: prnt.0ut: date: dd.mmm.yy
(menu code 1. 6. 9. 1) or
setup: prnt.oUt: date: mmm.dd.yy
(menu code 1. 6. 9. 2)
! No ISO/GLP-compliant record
is output if any of the following
settings are configured:
Setup: Prnt.Out: Print: Aut. W/o or
Aut.With (menu code 1. 6. 1. 3, 1. 6.
1. 4, ) or Format: 16 Char. (menu
code 1. 6. 6. 1)
Function Keys
Transfer header and first measured
value: press
> The header is included with the
first printout/data record.
To output header and reference
data automatically when an
application program is active:
press
Exit the application:
1) To send the GLP footer:
press
2) Quit application program:
press again
CLEAR
CLEAR
ENTER
PRINT
ISO/GLP-compliant Printout/Record
Features
You can have device information,
ID texts and date and time printed
before (GLP header) and after (GLP
footer) the values of a weighing
series. These parameters include:
GLP header:
Date
Time at beginning of measurement
Balance manufacturer
Balance model
Balance serial number
Software version number
Identification number of the
current sampling operation
GLP footer:
Date
Time at end of measurement
Field for operator signature
! Operating the Balance with
Printer:
Select the following settings on
the balance and on the printer:
Software handshake:
Setup: interf.: handshk.: sftware.
(menu code 1. 5. 4. 1)
Configuration
§ Setting menu codes for the printout
(see Configuration):
ISO/GLP-compliant printout or
record only for calibration/adjust-
ment:
Setup: Prnt.Out: GLP: Cal.-Adj.
(menu code 1. 6. 7. 2) or
ISO/GLP-compliant printout or
record always on:
Setup: Prnt.Out: GLP: Always on
(menu code 1. 6. 7. 3)