The ISO/GLP-compliant printout can contain the following lines:
-------------------- Dotted line
17-Aug-2005 10:15 Date/time (beginning of measurement)
DENVER Balance manufacturer
Mod. SI-8001 Model
Ser. no. 10105355 Balance serial number
Ver. no. 00-36-01 Software version
ID 2690 923 ID
-------------------- Dotted line
LID Measurement series no.
nRef 10 pcs Counting: reference sample quantity
wRef 21.14 g Counting: reference weight
Qnt + 567 pcs Counting result
-------------------- Dotted line
17-Aug-2005 10:20 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line
ISO/GLP-compliant printout for external calibration/adjustment:
-------------------- Dotted line
17-Aug-2005 10:30 Date/time (beginning of measurement)
DENVER Balance manufacturer
Mod. SI-8001 Model
Ser. no. 10105352 Balance serial number
Ver. no. 00-36-01 Software version
ID 2690 923 ID
-------------------- Dotted line
Cal. Ext. Test Calibration/adjustment mode
Set + 5000.0 g Calibration weight
Diff. + 0.2 g Difference determined in calibration
Cal. Ext. Complete Confirmation of completed calibration procedure
Diff. 0.0 g Difference from target following adjustment
-------------------- Dotted line
17-Aug-2005 10:32 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line