EasyManua.ls Logo

Denver Instrument Summit - Mass Unit Conversion Application; Features

Denver Instrument Summit
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Mass Unit Conversion
Purpose
With this application program you
can change the weight value dis-
played from the basic weight unit to
any of 4 application weight units (see
table on next page).
Preparation
§ Select the Unit application for
toggling weight units:
see chapter on Configuration
(Parameter Settings)
§ Set the following parameters:
Application program
unit Toggle wt, unit
dec.plcs. Display accuracy
o All All decimal places
minus 1 Reduced by 1 place
o = Factory setting
Features
Set the basic unit and display accu-
racy in the Setup menu: see Con-
figuration.
Set the application weight units
and display accuracies in the
Application menu.
These settings are stored in bat-
tery-backed memory.
The basic unit is active when the
balance is powered up.
Printout for Wt. Unit Toggling
+ 100.0 g : Weight with 16-character data output format
N + 0.22046 lb : Weight with 22-character data output format
13-Jan-2005 08:35 : Data output format for two-line printout:
N + 3.5275 ozt : Date/time and weight