41
Example: Determining the density of a solid sample.
Parameter settings:
Applic.: Density: Dec.Plcs: 1 Dec. Pl. (menu code 2. 9. 1. 2)
Step Key Display/Data output
(or instruction)
1.
Attach sample holder to suspension wire
2. Tare the balance t 0.0 g
3. Start application program
4. Confirm Air display
Air
?
5. Determine the weight
of the sample in air:
Place sample on the balance +
20.0 g
*
6. Store value for weight in air
7. Remove sample from the balance
water
?
8. Determine weight in liquid:
place sample in holder
9. Confirm water display
0.0 g
*
10. Immerse sample in liquid + 15.0 g
*
11. Store value for weight in
liquid, view result, and print +
4.0
o
*
Wa + 20.0 g
Wfl + 15.0 g
Rho 4.0 o
12. Delete result
13. Repeat as desired, starting from Step 3.