Internal Calibration/Adjustment (Only on SI models)
Set the following parameters:
Setup: Bal.Scal.: Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)
The built-in motorized calibration weight is applied and removed automatically
for internal calibration.
Step Key (or instruction) Display
1. Tare the balance
t 0.0 g
2. Start calibration
q Cal.Int.
The built-in weight is
Cal. run.
applied automatically
3. Calibration/adjustment executed
Cal.end
4. The built-in weight is removed 0.0 g