0065-0647-OM Rev E 7550 OWNER’S MANUAL
SCALE SETUP, CONT.
Settings Page 2
SETTINGS PAGE 2
OIML: No
Units: lb, in
Filter Mode: 0
Zero Tracking: 10
<<< Exit >>>
OIML:
Touching this key will toggle the OIML setting to Yes or No. NOTE: This setting should be
enabled for all EU models to comply with regulations.
Allowable settings for OIML are: Yes or No.
Units:
This is for reference only. The weighing/height units selection can ONLY be made when the
scale is turned on for the first time. At that time, a prompt will be displayed to select the
weighing/height units to use, pounds/inches (lb/in), or kilograms/centimeters (kg/cm). Once
the weighing/height units have been set, they cannot be changed.
Refer to the WEIGHING UNITS/HEIGHT MEASUREMENT section of this manual.
Filter Mode:
Pressing this key will open the weight filtering screen. This will allow you to set the amount
of digital filtering being applied to the scale. Using the numeric keys, enter a new setting,
and then press Save.
Allowable settings are: 0 = Off, 1 = Minimal, 2 = Moderate, and 3 = Maximum
Zero Tracking:
Touching this key will open the zero-tracking screen. This will allow you to set the number of
half (1/2) divisions that the scale will attempt to maintain zero. Using the numeric keys, enter
a new setting, and then press Save.
Allowable settings for Zero Tracking are 0 – 20.
Press >>> to advance to the Settings Page 3.
Press <<< to return to the Settings Page 1.
Press Exit to exit Setup and return to the Weight screen.