EasyManuals Logo

Detecto 7550 User Manual

Detecto 7550
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
0065-0647-OM Rev E 7550 OWNER’S MANUAL
24
SCALE SETUP, CONT.
Settings Page 4
SETTINGS PAGE 4
Decimal 1: 1
Decimal 2: 1
Motion Range: 3
Gravity: 0.000000
<<< Exit >>>
Decimal 1:
Touching this key will open the decimal entry screen. This will allow you to set the decimal
point precision for the first weight range. Using the numeric keys, enter a new setting, and
then press Save.
Allowable settings for Decimal 1 are 0 – 3.
Decimal 2:
Decimal 2 is only used when Capacity 2 is used. The Wheelchair Model selection does not
use Capacity 2 and Decimal 2. Note that even though Decimal 2 is not used, the factory
setting for Decimal 2 is set to 1.
If Decimal 2 is not 1, touch this key to open the Interval 2 screen. Using the numeric keys
enter a 1, and then press Save.
Motion Range:
Touching this key will open the motion range entry screen. This will allow you to set the
number of scale divisions of movement that will be allowed for a stable weight. Using the
numeric keys, enter a new setting, and then press Save.
Allowable settings for Motion Range are 0 – 20.
Gravity:
Gravity compensation accounts for latitudes and elevations that are different from where the
scale was calibrated. To calculate the setting for this parameter, use the gravitational
constant of the location where the scale was calibrated divided by the gravitational constant
of where the scale will be installed:
Gravitational Constant (Calibration location)
---------------------------------------------------------- = setting
Gravitational Constant (Operation location)
This should give you a setting close to 1 that you can enter in to compensate for variation in
gravity due to elevation/latitude.
Touching this key will open the gravity compensation entry screen. Using the numeric keys,
enter a new setting, and then press Save.
NOTE: Set Gravity to 0 (zero) to disable the gravity compensation feature.
Press >>> to advance to the Settings Page 5.
Press <<< to return to the Settings Page 3.
Press Exit to exit Setup and return to the Weight screen.

Table of Contents

Other manuals for Detecto 7550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Detecto 7550 and is the answer not in the manual?

Detecto 7550 Specifications

General IconGeneral
BrandDetecto
Model7550
CategoryScales
LanguageEnglish

Related product manuals