0065-0647-OM Rev E 7550 OWNER’S MANUAL
SCALE SETUP, CONT.
Settings Page 6
SETTINGS PAGE 6
Serial Type: Pass
USB: SMA
Print: Manual
<<< Exit >>>
Serial Type:
The Wheelchair Model selection does not use the Serial RS232 port. Note that even though
the Serial Type is not used, the factory setting for Serial Type is set to Pass (pass-through
printing).
If Serial Type is not Pass, touch this key to open the serial type selection screen. Using the
numeric keys enter a 0, and then press Save.
USB:
The Wheelchair Model selection does not use the USB port. Note that even though the USB
Port is not used, the factory setting for USB Port is set to SMA.
If USB Port is not SMA, touch this key to open the USB Port selection screen. Using the
numeric keys enter a 0, and then press Save.
Print:
Touching this key will toggle the Print setting to Auto or Manual for printing (outputting data)
when a stable weight has been obtained on the scale.
If “Auto” is selected, the scale will automatically print (output data) as soon as a stable
weight is obtained.
If “Manual” is selected, after a stable weight has been obtained, you must press the
key to print (output data).
Allowable settings for Print are: 0 = Manual, 1 = Auto
Press >>> to advance to the Settings Page 7.
Press <<< to return to the Settings Page 5.
Press Exit to exit Setup and return to the Weight screen.