0065-0647-OM Rev E 7550 OWNER’S MANUAL
34
CALIBRATE SCALE
1 = SMARTCAL, CONT.
8. The display will change to show “Working” for a few seconds and then change to show “Set
test weight over Cell 3”.
CALIBRATE SCALE
Set test weight
over Cell 3 and
press Save.
1 2 3
4 5 6
7 8 9
SAVE CANCEL
0 .
DEL
9. Place the calibrated test weight on the scale over Cell 3 and then press Save.
10. The display will change to show “Working” for a few seconds and then change to show
“Set test weight over Cell 4”.
CALIBRATE SCALE
Set test weight
over Cell 4 and
press Save.
1 2 3
4 5 6
7 8 9
SAVE CANCEL
0 .
DEL
11. Place the calibrated test weight on the scale over Cell 4 and then press Save.
12. The display will change to show “Working” for a few seconds and then return to the
Settings Page 5 screen.
13. Calibration is now complete.
14. Press Exit to exit Setup and return to the Weight screen.