EasyManuals Logo

Detecto solo Series User Manual

Detecto solo Series
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
0044-0391-0M Rev F y solo Series Owner’s Manual
16
KEYPAD FUNCTIONS, CONT.
This is the LOCK/RELEASE key. It is used to lock and unlock the display.
Pressing this key will cause the display to lock onto the weight and remain
displayed after the patient steps off the scale. Pressing the key again will unlock
the display and return it to zero.
NOTE: The scale will not respond to pressing the key unless the weight is
stable, and the STA annunciator is on.
This is the BMI key. It is used to first input the patient’s height, and then to
perform the Body Mass Index (BMI) calculation. Note that the scale will not
respond to pressing the key unless the weight is stable.
When using the solo height rod (or no height rod), pressing the key once allows
the operator to enter the height of the patient using the © or ª keys to increment
or decrement to the correct height. After entering the patient’s height, pressing it
again will display the Body Mass Index (BMI) calculation. Pressing it a third time
will return the display to the patient’s weight.
NOTE: While the patient is still standing on the scale, pressing the
key will
toggle between the BMI and weight display. Also, note that once the patient steps
off the scale, the display automatically returns to the weight only mode.
This is the UP ARROW and MOTHER/BABY key. It is used for two purposes.
When performing the Body Mass Index (BMI) calculation, it is used to increase
the value for the patient’s height.
When using the Mother/Baby Weighing feature, this key is used to determine the
weight of infants and toddlers held by an adult. This feature allows the adult to be
weighed first, and then the adult and infant/child weighed with only the weight of
the infant/child displayed.
This is the DOWN ARROW key. It is used when performing the Body Mass Index
(BMI) calculation to decrease the value for the patient’s height.
NOTE: During setup and calibration, the arrow keys are used to select setup and
calibration values.