0044-0391-0M Rev F y solo Series Owner’s Manual
28
DISPLAY MESSAGES
The solo Series Digital Clinical Scales are equipped with software that tests various portions
of the scale's circuitry and verifies proper operation. Should a problem be detected, a
message will be displayed alerting the operator to that condition. The following lists the
messages displayed and their meaning:
DISPLAY MEANING
99
This will be displayed for a few seconds when the scale cannot establish
zero, and then turn the scale off. Refer to the TROUBLESHOOTING
section.
!6;`
Displayed when the batteries are near the point they need to be replaced.
#$
Weight on scale exceeds 570 lb (260 kg).
TROUBLESHOOTING
Your scale has been designed to provide you with years of trouble-free operation. Despite
this, troubles sometimes happen. Before calling for service assistance you should make
some initial checks to verify that a problem does exist. The following describes several types
of symptoms along with suggested remedies.
PROBLEM POSSIBLE SOLUTIONS
The display
does not turn on
Battery Operation:
y Check if the batteries are installed and correctly.
y Are the batteries discharged? Remove the old batteries and replace
them with new ones.
AC Operation:
y Is the 9V AC/DC power adapter fully inserted into the wall receptacle?
y Check the wall receptacle for proper AC power. Try another electrical
appliance in the same receptacle. Does it work?
y Check the circuit breaker.
y Has there been a power failure?
Incorrect weight
displayed
Make sure the scale platform is not touching an adjacent object.
Have proper operation procedures been followed?
The display
shows 99
and shuts off
If the platform is not empty (the patient is on the scale) when powering
up, the display will show 99 and turn off. Remove any weight from the
platform (have the patient step off the scale) and power up the scale
again. If the scale continues to show 99 and turns off, consult your
scale service representative.
The display
shows a moving
small
6
If the display shows a moving small
q
when powering up, make sure the
scale is stable and wait for a moment until the display shows ..
If the scale continues to show a moving small
q
, please consult your
scale service representative.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Detecto solo Series and is the answer not in the manual?

Detecto solo Series Specifications

General IconGeneral
BrandDetecto
Modelsolo Series
CategoryScales
LanguageEnglish

Related product manuals