0044-0391-0M Rev F y solo Series Owner’s Manual
25
CALIBRATION
Your solo Series Digital Clinical Scale was calibrated at the factory and should not require
adjustment. If the scale should need re-calibration, the following describes the calibration
procedure. To maintain the instrument’s high degree of accuracy, a qualified technician
should perform this function.
To perform calibration, you will need either 400 pounds or 200 kilograms of certified test
weights, depending on the weighing unit (kilograms or pounds) of your scale. It is
recommended to use eight (8) single 50-pound or 25-kilogram test weights for calibration.
NOTE: Before starting calibration, verify that the test weights match the weighing
units of the scale. If they do not match, the weighing unit setting on the scale must
be changed.
Changing the Weighing Unit
IMPORTANT! If the OneWeigh feature has been enabled, the '0-; setting cannot
be accessed until the OneWeigh feature has been reset. To reset (disable) the
OneWeigh feature, contact Detecto Technical Support (e-mail tech@cardet.com or
call 1-866-254-8261) for the procedure to reset it.
1. To change the weighing unit, ensure OneWeigh is disabled, and then turn the scale off.
2. Press and hold the ¨0§ key and then press the key to turn the scale on.
3. The scale will enter setup mode with the display showing '")& and flash the lb or kg
annunciator for the current setting.
4. Release all the keys.
5. Use the © or ª keys to change the weighing unit to lb or kg, (to match your test weights)
and then press the key.
6. Press the key repeatedly until the scale automatically turns off.
7. The scale is now ready to enter calibration.
To Enter Calibration
1. With the scale off, press and hold the and keys, and then press the key.
2. The scale will beep twice, and then the display will turn on showing *;.
3. Release all the keys.
4. The scale is now ready to begin calibration.
NOTE: When a prompt and value displayed are acceptable, press the
key to save the setting and proceed to the next prompt. To change a setting,
press the © or ª keys to adjust the setting (or select a new value) and then
press the
key to save the new setting and advance to the next prompt.