Cleaner tank is
empty, or the
dil./cleaner sensor
is too sensitive (try
recalibrating the
sensors), or there
are lot of bubbles
in the cleaner tube
Replace the cleaner reagent
container with a filled one.
If this is still a problem with a filled
cleaner container, run Reagent
sensor calibration and check the
tubing and fittings of cleaner path.
Rinse tank is
empty, or the rinse
sensor is too
sensitive (try
recalibrating the
sensors), or there
are lot of bubbles
in the rinse tube
Replace the rinse reagent container
with a filled one.
If this is still a problem with a filled
rinse container, run Reagent sensor
calibration and check the tubing and
fittings of rinse path.
The sampler rotor was jammed
during sampling process!
Retry sampling?
Cap was not taken
off from the
sample tube, or
tube is not placed
correctly, or other
mechanical jam
Check the adapter and the tube.
Press Retry or Stop.
The sampler rotor was
jammed!
Mechanical
problems like
adapter is moved
out from the right
place
Check the adapter and the tube.
Press the button
Unrecoverable pneumatical
error occurred!\n
Device: %s\n
Error code: %s\n
Status: %x\n
Please turn off the instrument
and turn it on again!
Fatal error in
mechanical
subsystem.
Possibly
mechanical jam.
Check the proper operation of the
motor modules (home/end opto
switches, ribbon cables and the
motor moving).
Fatal HGB error.
Power off the system!
HGB channel did
not give a ready
signal!
Check the connections between
instrument and printer and check
printer setup in Printer Settings