6.5.1. Error code list
%s: filename concerned, %d: error type concerned, %u: error specific string
Fatal system or
program error!
Check the COMB card and Dimm-PC
module for proper insertion.
Try to make a software upgrade.
Try to replace the DimmPC.
Fatal system or
program error!
The same as in case of error code
1000.
Data file %s is corrupt, new
will be created, all stored
measure data will be lost
Stored data file is
corrupt or missing!
Disk or software
error.
New stored data file will be
created automatically (by a User
confirmation).
Error %d on indexing file %s
Fatal system or
program error!
The same as in case of error code
1000.
Pressure error.\n
Power off the system
Cannot make the
measuring
vacuum!!
Check the pump and the tubes,
fittings and valves around the
puffer reservoir.
Diluent tank is
empty, or the
dil./clean sensor is
too sensitive (try
recalibrating the
sensors), or there
are lot of bubbles in
the diluent tube
Replace the diluent reagent
container with a filled one.
If this is still a problem with a filled
diluent container, run Reagent
sensor calibration and check the
tubing and fittings of diluent path.
Lyse tank is empty,
or the lyse sensor
is too sensitive (try
recalibrating the
sensors), or there
are lot of bubbles in
the lyse tube
Replace the lyse reagent
container with a filled one.
If this is still a problem with a filled
lyse reagent container, run
Reagent sensor calibration and
check the tubing and fittings of
lyse path.
The measuring
vacuum has been
dramatically
decreasing during
the measurement
or absolutely lost!
Leakage in the pneumatic system
(check the pump, fittings, tubing,
pressure sensor, puffer reservoir
and valves).
Pneumatical error occurred!
Device: %s
Error code: %s
Status: %x
Retry?
Fatal error in
mechanical
subsystem.
If frequently happens this error
check the proper operation of the
motor modules (home/end opto
switches, ribbon cables and
moving of motors).