32 Scale Link 2 Technical Manual D4204-EN
HOLD Performs a HOLD function, pull line low again to display “held” weight. Press ON to cancel HOLD
and display weight change.
NETGRS Performs a NET/GROSS function.
M+ Performs a M+ function, when line is pulled low the displayed weight will be stored to memory. This
will be followed by RM and display the memory total.
ZERO Performs a zero/balance
MIXCTR Enables the “Rotation Counter” on equipped indicators.
INGRED Used with EZ3410/TMR3610/TMR4610. If a recipe is loaded, will perform an ingredient advance. If
no recipe is loaded a TARE will be performed.
RECHCK Rechecks the weight when using the “lock-on” weigh method with Stock Weigh indicators.
SEEDTD Used to advance Seed Tender scale to next seed dispense operation; re-enters preset.
ST STP Performs a START/STOP function when using Harvest Tracker indicators.
FSWTCH Used with RPM START/STOP CONTROL (RSSCTL - D.A.N. 6401) to perform a START/STOP
function with the use switch. See Feedbox Start/Stop Polarity (FS POL D.A.N 6411) for switch
polarity.
TR HLD When enabled will display and remain on the GROSS weight if T/R button is held for three (3)
seconds. If not enabled, the GROSS weight will be momentarily displayed if T/R button is held for
three (3) seconds.
RI1MSG (D.A.N 1402) REMOTE INPUT 1 SWITCH MESSAGE
Used to edit the remote input 1 message that is displayed for an ‘OPEN’ or ‘CLOSED’ condition on the remote input
line of the power cord. The default message is ‘OPEN
R1STAT (D.A.N 1403) REMOTE INPUT 1 SWITCH STATE
Determines if the Remote Switch Message or Alarm Output lamp is displayed/illuminated for an ‘OPEN’ or
‘CLOSED’ condition on the remote input line of the power cord. The default setting is ‘CLOSED’.
R1TIME (D.A.N. 1404) REMOTE INPUT 1 SWITCH TIME
Used to set how often the Remote Switch Message 1 is displayed for an ‘OPEN’ or ‘CLOSED’ condition on the
remote input line of the power cord. For the default value of 2, the message is displayed every 2 seconds. The
maximum setting is 9 seconds.
RMINP2 (D.A.N. 1411) Remote Input 2 (Remote port Pin 3/TR)
When line is pulled low the same functions as remote input 1 can be performed.
(See Remote Input 1 (RMINP1 D.A.N 1401) for available settings.
RI2MSG (D.A.N 1412) REMOTE INPUT 2 SWITCH MESSAGE
Used to edit the remote input 2 message that is displayed for an ‘OPEN’ or ‘CLOSED’ condition on the remote input
line of the power cord. The default message is ‘OPEN