Close the rotor lid, firmly tighten clockwise the lid to the rotor and ensure is in
position. Close the door and then start running.
The method of removing the rotor is as same as the above mentioned by turning
the locking nut counterclockwise.
8.1.2Set the operation parameters
Press the key to select required parameters. The parameter can be modified
when the parameter is flashing. Clockwise rotate the parameter key to increase
parameter value; counter-clockwise rotate the parameter key to decrease
parameter value. Parameter key rotate faster, parameter value increase faster.
The minimum speed increment is 100 rpm, the minimum time increment is 1 second.
1Set the speed
Press the key until the speed rpm is displayed.
When the speed is selected, the speed symbol will flash the speed value.
The minimum speed value you can set 300rpm, the minimum increment is
100rpm.
Rotate parameter key clockwise to increase speed value, rotate the
parameter key anti-clockwise to decrease speed value.
You can speed-up set the speed value by rotating parameter key quickly.